Experimental Work:
There were three samples in the experiment which were indicated by “Sample A”, “Sample B” and “Sample C”. Sample C was provided to me for hardness measurement. The apparatus required for this experiment is given below;
- Vicker Hardness Testing Machine
- Sand paper ( If sample surface is not clean)
- Sample
Procedure:
The procedure of the experiment was involved following steps;
- The sample C was provided me to perform the hardness test; the first job was to clean the surface of the sample by sand paper.
- After cleaning the surface of specimen the, the sample was placed upon the table.
- The load could be varied and selected from the screen on the vicker hardness tester.
- The specimen was then brought near to the diamond indenter.
- The microscope was turned out of the position while the indentation is being made.
- The load was then applied on the specimen.
- After producing indent the microscope was focused on the indent to study the indent.
- Proper magnification is important for accurate measurements.
Calculations:
Indenter used = Highly polished pointed square base pyramidal diamond with face angles of 136.
Load applied = 9.807N
Lens used = 4x
Time = 10 sec
L-1 (Horizental Diagonal) = 104.28µm
L-2 (Vertical Diagonal) = 96.90µm
HV = 183
HRC= 6.7